Diagnostic Accuracy of a Handheld Ultrasound vs a Cart-based Model: A Randomized Clinical Trial.
Gibbons, Ryan C. ; Jaeger, Daniel J. ; Berger, Matthew ; Magee, Mark ; Shaffer, Claire ; Costantino, Thomas G.
Gibbons, Ryan C.
Jaeger, Daniel J.
Berger, Matthew
Magee, Mark
Shaffer, Claire
Costantino, Thomas G.
Citations
Altmetric:
Genre
Journal article
Date
2024-02-09
Advisor
Committee member
Group
Department
Emergency Medicine
Subject
Permanent link to this record
Collections
Files
Research Projects
Organizational Units
Journal Issue
DOI
https://doi.org/10.5811/westjem.17822
Abstract
Introduction: Numerous studies have demonstrated the accuracy of point-of-care ultrasound (POCUS). Portable, handheld devices have expanded the clinical scope of POCUS at a fraction of the cost of traditional, cart-based models. There is a paucity of data assessing the diagnostic accuracy of portable devices. Our objective in this study was to compare the diagnostic accuracy of a portable device with a cart-based model.
Methods: This was an institutional review board-approved, observational, prospective, randomized clinical trial (NCT05196776) of a convenience sample of adult patients who presented to a university-based health system. Patients who required a cardiac, lung, renal, aorta, or biliary POCUS were randomized to a portable device or to a cart-based model. We hypothesized that the cart-based model would have a 90% diagnostic accuracy vs 70% for the handheld device. To detect a 20% difference, the sample size was calculated to be 98, with 49 patients randomized to each arm. We used standard 2x2 tables to calculate test characteristics with 95% confidence intervals (CI).
Results: A total of 110 patients were enrolled, with 56 patients randomized to the cart-based model and 54 to the handheld device. The sensitivity, specificity, and diagnostic accuracy of the cart-based vs handheld were 77.8% (40-97.2) vs 92.9% (66.1-99.8), 91.5% (79.6-97.6) vs 92.3% (79.1-98.4%), and 89.3% (78.1-96) vs 92.5% (81.8-97.9), respectively.
Conclusion: The diagnostic accuracy of a portable, handheld device is similar to that of a cart-based model.
Description
Citation
Gibbons RC, Jaeger DJ, Berger M, Magee M, Shaffer C, Costantino TG. Diagnostic Accuracy of a Handheld Ultrasound vs a Cart-based Model: A Randomized Clinical Trial. West J Emerg Med. 2024 Mar;25(2):268-274. doi: 10.5811/westjem.17822. PMID: 38596929; PMCID: PMC11000544.
Citation to related work
University of California
Has part
The Western Journal of Emergency Medicine, Vol. 25, Iss. 2