Genre
Pre-printDate
2018-12-10Author
Lu, BTorchinsky, DH
Permanent link to this record
http://hdl.handle.net/20.500.12613/4365
Metadata
Show full item recordDOI
10.1364/OE.26.033192Abstract
© 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement We describe a novel scheme of detecting rotational anisotropy-second harmonic generation (RA-SHG) signals using a lock-in amplifier referenced to a fast scanning RA-SHG apparatus. The method directly measures the nth harmonics of the scanning frequency corresponding to SHG signal components of Cn symmetry that appear in a Fourier series expansion of a general RA-SHG signal. GaAs was used as a test sample allowing comparison of point-by-point averaging with the lock-in based method. When divided by the C∞ signal component, the lock-in detected data allowed for both self-referenced determination of ratios of Cn components of up to 1 part in 104 and significantly more sensitive measurement of the relative amount of different Cn components when compared with conventional methods.Citation to related work
The Optical SocietyHas part
Optics ExpressADA compliance
For Americans with Disabilities Act (ADA) accommodation, including help with reading this content, please contact scholarshare@temple.eduae974a485f413a2113503eed53cd6c53
http://dx.doi.org/10.34944/dspace/4347